Chapter 6
Gated Operation
71
Figure 42. Single Shot Result: Fluorescence Spot, 100 ns Width, 10 ns Delay, Binned Vertically
Procedure for Setting up and Performing a Swept Gate Experiment
(Variable Width, Variable Delay):
The sequence of steps for a Swept Gate experiment with variable width and delay is
similar to the Swept Gate experiment with fixed width and variable delay but with the
following exception:
In addition to entering differing start and end values for the Gate Delay, you would
also enter differing start and end values for the Gate Width.
Procedure for Setting up and Performing a Static Gate Experiment
(Fixed Width, Fixed Delay)
The sequence of steps for a Static Gate experiment is similar to the Swept Gate
experiment but with the following exceptions:
The Active Mode, selected in Step 100, would be
Repetitive
.
The Gate Width and Gate Delay values would be entered in the dialog shown below.
Note that the minimum gate delay from T0 to optical gate is ~10 ns:
Figure 43. Repetitive Gating Setup dialog box
Summary of Contents for PI-MAX 3 System
Page 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Page 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Page 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Page 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...