46
PI-MAX
®
3 System Manual
Version 1.A
Partial Frame ROI Binning
When setting up a partial frame ROI (regardless of the array type – full frame or interline),
keep in mind that for the PI-MAX3 the following constraint applies:
the number of pixels in
the serial (horizontal) direction must be evenly
divisible by 4, even after binning.
The software
may refer to the horizontal as X or Wavelength
depending on the application.
Examples:
These examples include partial frame
ROIs with and without binning. The terminology is
based on the WinView tab page shown to the right.
X Start to End
= 200 pixels, no grouping (binning).
Since 200/4=50, this is a valid ROI setup.
X Start to End
= 200 pixels and grouping (binning)
is by 8. The resulting number of super pixels is
25. Since 25/4=6.25, this is not a valid ROI
setting for the horizontal direction.
X Start to End
= 240 pixels, no grouping (binning).
Since 240/4=60 this is a valid ROI setup.
X Start to End
= 240 pixels and grouping
(binning) is by 3. The resulting number of
super pixels is 80. Since 80/4=20, this is a
valid ROI setup.
X Start to End
= 240 pixels and grouping
(binning) is by 16. The resulting number of super pixels is 15. Since 15/4=3.75, this is not a
valid ROI setting for the horizontal direction.
Figure 15. Partial Frame 2 × 2 Binning for Interline CCD - Single Port Readout (ROI is not
centered and symmetrical horizontally)
Figure 14. Experiment Setup|ROI Setup tab
8 is Not Valid
For X Start-End
of 200
Summary of Contents for PI-MAX 3 System
Page 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Page 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Page 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Page 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...