CAL: display status of control board calibration. If this shows “FAIL”, it has lost
the calibration data and you need to call Tech Support.
5.2.3. Inverter status
No function for this power supply.
5.2.4. Auxiliary status
W/IN TEMP: Displays water inlet temperature as centigrade (
℃
).
5.2.5. Date/Time display and Setting
Display or setting date & time
To set date and time, move cursor to the item use UP (▲), DOWN (▼), LEFT (◄)
and RIGHT (►) key and enter value with numeric keys and press “ENT” key to
change and move cursor to next item.
Press “SAVE” key to save changes and return to FUNCTION menu.
5.2.6. History
Display history events of Power supply.
Stores and displays the numbers of failed events that occurred in the heating
process.
TURN ON: Stores time of power on of power supply. (hour, minute)
HEATING: Stores time of Heat on. (hour, minute)
PRESS FAIL: Stores failure of differential pressure of cooling water.
TEMP FAIL: Stores if exceeding temperature of cooling water.
INV FAIL: Not activated for this power supply.
SW/TH FAIL: Number of swing thru fail count.
SHORT COIL: Number of resonant trips.
INV.C TRIP: Number of inverter current trips.
DOOR OPEN: Number of door opens.
E-STOP: Number of emergency stops.
AUX-INT: Number of auxiliary interlocks.
5.2.7. Version
Display firmware and PLD version of Control board and display board.
(CONT = control board, DISP = display board, S000 = firmware version, P000 = PLD
version)
19