Sens 2. Prim. Meas.
: item 6 on the home screen.
Sens 2. Sec. Meas.
: item 13 on the home screen.
I1: Source signal 1
desig.
------------ : no designation
Water height
Defined by User
: in this case, an additional line will appear as soon as this
option is confirmed
I1: User desig.
A free text field for entering the user's designation (maximum of
13 characters).
I1: Source signal 1
stability delta
Can be set from 0 to 50 % (the default is 1 %) and is used to define when
the arrow on the home screen will indicate that the measurement is stable if
the parameter measured has been set as the primary parameter for sensor
1 or 2 (refer to items 2 and 5 in Table 2)
I1: Source signal 2
Inactive
Active
: source signal 2 is calculated from source signal 1 by a conversion
program written in a file stored on the USB stick plugged into the USB port.
The name of the file must be of the form "CONFANA2.csv" and must
contain the following information:
Line 1
of the csv file: a description (not more than 16 characters) of the
calculated source signal.
Line 2
:
a blank line containing no characters.
Lines 3 to 32
: a number of conversion points in the form of portions of a
straight line Source signal 2 = f(Source signal 1).
To activate the conversion, plug the stick into the ACTEON 5000's USB port
(refer to item 6 of diagram 3) and wait for the small green LED located
above the USB port to stop flashing.
Then navigate to menu I2: Source signal 2 and select the "Active" option.
If Source signal 2 is Activated
I1: Source signal 2
display
Not displayed
Sens 1. Prim. Meas.
: the measurement shall be displayed in the space
allocated to the primary parameter for sensor 1 (item 3 on the home screen)
Sens 1. Sec. Meas
: item 9 on the home screen.
Sens 2. Prim. Meas
: item 6 on the home screen
Sens 2. Sec. Meas
: item 13 on the home screen.
I1: Source signal 2
stability delta
Can be set from 0 to 50 % (the default is 1 %) and is used to define when
the arrow on the home screen will indicate that the measurement is stable if
the parameter measured has been set as the primary parameter for sensor
1 or 2 (refer to items 2 and 5 in Table 2)
Setting up Input 2
Input 2
Programming of Analog Input No.2:
Inactive
(no programmed action)
Current
: if a sensor is connected which generates a signal of 0/4-20 mA
Voltage
: if a sensor is connected which operates within a 0-10 V range
I2: Signal min
For a
Current
input: can be set to a value between 0 and 20 mA
For a
Voltage
input: can be set to a value between 0 and 10V
I2: Signal max
For a
Current
input: can be set to a value between the minimum value set
up in the "Signal min" line and 20 mA.