5.2 Installation of digital sensors.
When sensors are connected to the transmitter for the first time they must be installed by running a
SCAN (to scan the addresses from 1 to 243).
As soon as the transmitter detects a sensor, it displays its address, its serial number (which is also
engraved on the body of the sensor) and a description of the sensor.
In the example above, pH and oxygen sensors have been detected: the pH sensor is at address 24,
the sensor's serial number is SN-PPHRA-0762 and the description of the sensor is
pH/redox/Temperature Ponsel Mesure. The OPTOD sensor is at address 10, the serial number is SN-
PODOA-4032 and its description is OPTOD/Temperature Ponsel Mesure.
The operator can stop the SCAN operation at any time by pressing on the
icon
.
If no sensor is detected, or 2 sensors have the same address (an address conflict) or a communication
error is detected, then the device displays a warning message accompanied by the
icon.
Pressing on the icon accesses the addressing menu (refer to section 6.2.1 for more details
about the addressing function) where the operator can change the sensor's address in the event of an
address conflict (i.e. if 2 sensors have the same address).
5.3 Selection of the parameters.
Once ACTEON has detected the sensors, the parameters measured by the sensors must be set up.
For each sensor, a primary and a secondary parameter can be selected.
To access the measurement settings menu, follow the sequence shown below from the home screen:
To set up the primary and secondary parameters for sensor 1, select the corresponding icon.
To select a different line, press on the
and
icons.
Open the set-up line of interest using the Confirm icon,
.