From the "Numeric input settings" window, select the
icon then select the sensor to be
calibrated.
This window is used to calibrate the parameters measured by the selected sensor, or to set the
external compensation data.
Icon
Functionality
Used to access the sensor's calibration menu
This menu is used to set up the external data used for compensation purposes
The
external compensation data
is detailed in the table below:
Set-up line
Functionality
Atmo. pressure
Atmospheric pressure compensation is applied to the OPTOD sensor
(measurement of oxygen levels using optical technology). The atmospheric
pressure sensor is built into the ACTEON 5000.
The first line is used to activate the compensation and to access a second line
which is used to adjust this parameter (possible values: 0 to 2,000 hPa).
Salinity
Salinity compensation can be applied to the Oxygen in mg/L parameter
measured by the OPTOD oxygen sensor.
The first line is used to activate the compensation and to access a second line
which is used to adjust this parameter (possible values: 0 to 85.00 g/kg).
Temperature
Each sensor is equipped with its own temperature sensor, and the temperature
recorded is considered if temperature compensation is required (for the pH,
oxygen in mg/L and conductivity parameters). However, it is also possible to
enter a fixed external compensation value.
The first line is used to activate the compensation and to access a second line
which is used to adjust this parameter (possible values: 0.00 to 40.00 °C).
Alpha
This coefficient can be set for the Non-compensated conductivity parameter
measured by the CTZN sensor (possible values: 0.0 to 6.0 %/°C) in order to
activate linear compensation.
To select the line above or below, or to increase or reduce a value, use the
and
icons.
The
icon confirms the value set for the external compensation data. Any changes made become
effective when the "Communication completed successfully" message appears.
To access the "Calibration settings" menu from the "Calibration: sensor 1 (or 2)" window, select the
"Calibration" icon.