18
Version 2.1
Update: April 2016
PONSEL Mesure
– Aqualabo Contrôle
35 Rue Michel Marion
– 56850 CAUDAN, FRANCE
Tel.: +33(0)2 97 89 25 30
– Fax: +33(0)2 97 76 55 72
RS485 - White
V+ sensor power supply: Red
A
12V
A
12V
4
– ANALOG
OUTPUT
Two analog outputs
0-20 mA or 20-0 mA or 4-20 mA or 20-4
mA
Output 1: -
Output 1: +
Output 2: -
Output 2: +
GND
SA1
GND
SA2
GND
SA1
GND
SA2
5
– RELAY
OUTPUT
2 relay outputs
Output 1: Release (default is NC)
Output 1: Switch
Output 1: Operate (default is NO)
Output 2: NC
Output 2: Switch
Output 2: NO
R1
C1
T1
R2
C2
T2
R1
C1
T1
R2
C2
T2
6
– USB
USB port
For downloading data
USB
7
– Ethernet
Ethernet option
ETH
8
– 485 OUTPUT
One RS485 output
RS485 (-)
RS485 (+)
G
B
A
GND
B
A
9 - ANALOG
INPUT
Two analog inputs
Analog input 2
Input 2: -
Input 2: +
Analog input 1
Input 1: -
Input 1: +
Jumpers: Selection jumpers
Voltage or Current
GND
EA2
GND
EA1
U
I
GND
EA2
GND
EA1
U
I
A - RESET
RESET button
RESET
/
B- WAKE
A button which should not be activated
WAKE
/
Table 1: Description of the electrical connections