6.2.7 Configuring the Recording output.
Set-up line
Functionality
Recording
Inactive:
the recording function is not activated
Active
: this option activates the recording function
Measurement
variation
Possible values: 0.5 to 50 %. The measurement will not be recorded if its
variation (as a %) from the last recorded value does not meet this
requirement.
Interval
The interval between consecutive recordings can be set from 1 to 120
minutes
USB export
Inactive
Active
: after connecting a USB stick, the data recorded on the ACTEON
5000 is copied to the latter.
To download the recorded data, plug the stick into the ACTEON 5000's USB
port (refer to item 6 of diagram 3) and wait for the small green LED located
above the USB port to stop flashing.
Navigate then to the "USB export" menu and select the "Active" option.