2.2.3 EHAN sensor: Redox and temperature.
The PONSEL sensor incorporates a reference electrode, used for ORP measurements, such as Ag /
AgCl in saturated KCl electrolyte plasticized "PLASTOGEL" ®.
The electrolyte "PLASTOGEL" ® communicates directly with the external environment without
interposition of capillary or porous. So there is no risk of clogging or defusing the reference.
The measuring electrode is in platinum (3,5mm2) presented in sealed ring on a glass rod and is for in-
situ measurements continuously
Temperature
: measures via CTN.
ORP Measure
Principle of ORP measure
Combination Electrode (ORP/reference) platinium
ring, Reference Ag/AgCI. Gelled electrolyte (KCI)
Range of measures
- 1000.0 to + 1000.0 mV
Resolution
+/- 0.1 mV
Precision
+/- 10 mV
Answer time
< 90 s
Temperature measurement
Principle of measure T°C
CTN
Temperature
0,00 °C to + 50.00°C
Resolution
0.01 °C
Precision
+/- 0.5 °C
T90
< 300 s
Temperature of storage
0°C to + 60°C
Protection scale
IP 68
Interface signal
Modbus RS-485 standard and SDI-12 in option
Refresh rate measurement
Maximum < 1 second
Sensor power
5 to 12 volts
Consumption
Standby : 25 µA
Average RS485 (1 measure/second) : 20 mA
Pulse current : 500 mA
Heating time : 100 mS
Sensor
Dimensions of sensor mounted
Mounted sensor length: gland (262 mm) not
included ;
Length with gland : 324 mm.
Weight
350 g ( cable)
Material in contact with the environment
PVC, POM-C,platinum, Polyurethane
Maximum Pression
5 bars
Cable/ connection
9-wire shielded conductor, uncoated-strand
polyurethane sleeve