3D / SmartPan
•
3D s / 3D Classic: 528 mm (20.7 in.)
•
3D Plus or 3D Mid with X-ray tube D-054SB /
3D Max with X-ray tube D-067SB: 600 mm
(23.6 in.)
•
3D Plus / 3D Mid / 3D Max with X-ray tube
D-059SBR or SXR 130-10-0.5 SC: 632 mm
(23.8 in.)
Magnification
Pan
•
2D / 3D s / 3D Classic: 1.2-1.5
•
2D tomography: 1.5
•
3D Plus / 3D Mid: 1.4
SmartPan
•
3D s / 3D Classic: 1.27
•
3D Plus / 3D Mid / 3D Max: 1.4
Ceph
1.13
3D
•
3D s / 3D Classic: 1.58
•
3D Plus / 3D Mid / 3D Max: 1.38, 1.40, 1.42,
1.43 or 1.8
Duty cycle for height adjustment
25 s ON / 400 s OFF
Line voltage
100 - 220 V~ / 50 - 60 Hz
230 - 240 V~ / 50 Hz
Line current
8 - 17 A
Line harmonics
Cos better than 0.9
Max. permissible apparent impedance of supply mains
0.5 Ω (100 VAC)
Max. continuous heat dissipation
250 W
Internal fuse(s)
User replaceable
100 - 220 V~ / 16A FF H 500 V
230 - 240 V~ / 8A FF H 500 V
Type
195100 ELU
External fuse(s)
100 - 220 V ~ / 16A min. - 20A max. T 250 V
230 - 240 V ~ / 10A min. - 20A max. T 250 V
Battery
Lithium battery: CR2032, Varta / Panasonic
Max. weight
2D / 3D s / 3D Classic
119 kg (262 lb)
3D Plus / 3D Mid
141 kg (311 lb)
17 Technical data for Planmeca ProMax product family
User's manual
Planmeca ProMax 97