4. In the Planmeca Romexis program, enter a name for this material and
select OK.
NOTE
The calibration exposure values are automatically included at the
beginning of the name.
10.2 Taking an exposure of an impression or plaster cast
NOTE
The X-ray unit has to be calibrated for each new material that is used. Refer
to the previous section for details.
NOTE
FOR IMPRESSIONS:
Do not use impression trays made of metal.
10.2.1 Selecting settings
1. Remove any patient supports attached to the X-ray unit patient support
table (1). Insert the polystyrene disc (2) provided (part number
10030330) into the adapter (3) if it is not already in place.
10 3D model exposure
64 Planmeca ProMax
User's manual