Select this button to move the C-arm to the back, away from the patient
positioning area. This full view position allows you to monitor and adjust
the patient’s position freely from all directions.
•
Select this button to position the C-arm around the patient support table.
This is the traditional closed patient entry position.
NOTE
If needed, the full view position (top button) can be disabled (Settings > User
> 1300 Operational settings > 1330 Patient positioning > Side entry OFF).
This might be necessary if there is no space for the C-arm to move back.
8.4.2 Positioning patient’s head
1. Guide the patient to the X-ray unit. The patient can sit or stand during the
exposure.
NOTE
We recommend that you image patients with poor health in a sitting
position.
2. Adjust the X-ray unit to suit the height of the patient. To do this, press
either of the height adjusting buttons until the chin cup is approximately
level with the patient’s lower jaw.
3. Ask the patient to grasp the patient handles.
4. Check that the patient’s head is firmly positioned in the head support.
•
You can adjust the head support by turning the adjusting knob at the
top.
•
You can use fastening straps for additional head support if needed.
Refer to section "Attaching patient supports" on page 27 for details.
8.5 Selecting exposure values
To go to the next screen, select:
8 3D patient exposure
User's manual
Planmeca ProMax 35