Factory presets for image resolution High Definition (HD)
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
100
7.1
3.6
Small adult (S)
120
6.3
3.2
Medium-sized adult (M)
120
8
4
Large adult (L)
120
8
5
Extra large adult (XL)
120
8
6.3
X-ray units with 120 kV X-ray tube (SXR 130-10-0.5 SC)
Child (XS)
100
7.1
3.6
Small adult (S)
120
6.3
3.2
Medium-sized adult (M)
120
8
4
Large adult (L)
120
10
5
Extra large adult (XL)
120
11
6.3
Factory presets for image resolutions Hi Res and Endo
Patient size
kV value
mA value
mA value with
ULD
X-ray units with 90 kV X-ray tube
Child (XS)
90
8
5.6
Small adult (S)
90
10
7.1
Medium-sized adult (M)
90
12.5
9
Large adult (L)
90
14
11
Extra large adult (XL)
90
14
14
X-ray units with 120 kV X-ray tube (D-059SBR)
Child (XS)
100
9
4.5
Small adult (S)
120
8
4
Medium-sized adult (M)
120
8
5
Large adult (L)
120
8
6.3
Extra large adult (XL)
120
8
8
X-ray units with 120 kV X-ray tube (SXR 130-10-0.5 SC)
Child (XS)
100
9
4.5
Small adult (S)
120
8
4
Medium-sized adult (M)
120
10
5
Large adult (L)
120
11
6.3
Extra large adult (XL)
120
11
8
If you need to adjust the preset exposure values for this exposure:
1. Select the kV / mA field.
2. Use the minus or plus buttons to set the exposure values you wish to
use. To improve the image contrast, reduce the kV value. To reduce the
radiation dose, reduce the mA value.
3. Select the green check mark button.
8 3D patient exposure
40 Planmeca ProMax
User's manual