12 Help messages
The X-ray unit incorporates a self-checking feature that monitors the
operation of the unit. If the system detects an operating error, a help
message (e.g. H101) appears on the touch screen.
The X-ray unit will not accept any commands from the user until the help
message is cleared from the touch screen. Clear the message by touching
the green check mark.
The following list shows, in numerical order, all the help messages that can
appear.
Code
Explanation
Comments
H101
Exposure switch
The exposure button was released
before the end of the exposure.
Guide the patient away from
the X-ray unit before moving
the C-arm.
Press and hold down the
exposure button for the entire
duration of the exposure.
H102
The exposure button is stuck or the
cable is short circuited.
Release the exposure button.
Contact your service
technician if you need to
replace the exposure switch.
H103
ProTouch desktop
application
The ProTouch desktop application
needs to be closed.
Restart the application.
H105
Emergency stop
button
The emergency stop button has
been activated.
All movements of the X-ray
unit are blocked and no
radiation is generated.
Guide the patient away from
the X-ray unit. Then release
the emergency stop button to
resume normal operation.
H110
Device
Maintenance
The date for your next scheduled
device maintenance is due.
Please contact your service
technician to arrange a
service visit.
H112
Collimator
Check panoramic X-collimator
position setting.
Check collimator settings.
H113
Check panoramic X-collimator width
setting.
Check collimator settings.
H114
DEC calibration initialization failed.
Check the sensor connection.
H115
DEC
DEC is receiving too much radiation. Change the exposure values.
H116
DEC is receiving too little radiation.
Change the exposure values.
H127
CEPH
Cephalometric DEC exposure too
long.
H130
Patient safety area Patient safety area violation
detected.
Check the scan and layer
settings.
H142
Height movement
Height movement is not possible
because one (or more) of the
positioning control buttons or the
positioning joystick is stuck.
Clear any obstruction before
moving the column again.
12 Help messages
User's manual
Planmeca ProMax 83