Chapter H - PARTS REPLACEMENT & REPAIR
Planmeca Proline EC X-ray unit H-3
MOTORS
Technical Manual
2.2 Rotating motor & gear
a)
Remove the cover of rotating part.
b)
Remove the Generator processor PCB from its holders.
c)
Remove the loading spring.
d)
Remove the connector.
e)
Remove the phillips screw from the end of the axle.
f)
Remove the motor/gear package.
g)
Remove the spring nipple and plastic bushing from the gear.
Changing motor to gear
a)
Slack the stop screw of the rubber clutch
b)
Remove the nuts going trough the rubber plate (4 pcs)
c)
Remove the motor.
d)
Detach the rubber clutch from the motor axle.
e)
2.3 Cassette motor/gear
a)
Remove the outer secondary slot and back cover of the cassette head.
b)
Remove the loading spring of the cassette motor.
c)
Remove the connector.
d)
Remove the attaching screw of the cassette motors attaching shaft.
e)
Remove the motor.
f)
Remove the spring nipple and attaching axle with bearings.
2.4 Positioning motors
a)
Remove the cover of lower shelf.
b)
Remove all the connectors.
c)
Remove the handles.
d)
Remove the attaching bolts of the positioning mechanism.
e)
Remove the attaching mechanism.
Patient positioning mechanism motor
a)
Remove the panduit beside the patient positioning mechanism motor
b)
Remove the patient positioning mechanism motor with rubber plate.
c)
Remove the rubber plate and cogged wheel from the motor.
Summary of Contents for Proline EC ceph
Page 2: ......
Page 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Page 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Page 208: ......
Page 209: ......