Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline EC X-ray unit E-65
CEPHALOSTAT ADJUSTMENT
Technical Manual
Rotate the second primary slot with the two screws located on the right side of the slot until
the slot is correctly positioned (use the 2.5mm allen key).
Figure 117
Second primary slot too far to the right or left
Loosen the attachment screws of second primary slot with the 4mm allen key and move the
slot in horizontal direction.
Figure 118
NOTE
After checking and adjusting the second primary slot position the cephalostat sensor
head must be calibrated according to the instructions given in section “Calibrating the
cephalostat sensor head” on page E-52.