Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-50 Planmeca Proline EC X-ray unit
CEPHALOSTAT ADJUSTMENT
Technical Manual
To check the beam position in vertical direction place the beam alignment tool to the sensor
alignment tool to vertical position as shown on the Fig. 88 below.
Figure 88
Stand behind the tubehead and protect yourself from radiation. Press and hold down the
exposure button. The image of the radiation beam will appear on the beam alignment tool.
The radiation beam must reach the lower edge of the rectangle marked on the beam align-
ment tool. If it does not refer to “Radiation beam adjustment, Planmeca Proline EC Pan/Ceph
X-ray” on page E-15. The adjustment is performed in the same way as when adjusting the
standard panoramic slot.
Figure 89
Move the beam alignment tool to the upper position on the sensor alignment tool as shown on
the Fig. 89.
The radiation beam must reach the upper edge of the rectangle marked on the beam align-
ment tool (see figure above). If it does not refer to “Radiation beam adjustment, Planmeca
Proline EC Pan/Ceph X-ray” on page E-15. The adjustment is performed in the same way as
when adjusting the standard panoramic slot.
When all the slots have been checked, and if necessary adjusted, exit the service mode by
pressing the ceph key.
D2adj_1pr
im.eps
Beam alignment
Beam alignment
tool in upper
position
tool in lower
position
Summary of Contents for Proline EC ceph
Page 2: ......
Page 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Page 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Page 208: ......
Page 209: ......