Chapter G - TROUBLESHOOTING
G-6 Planmeca Proline EC X-ray unit
DISTURBANCES IN CASSETTE MOVEMENT
Technical Manual
5.3 Trembles during movement
5.4 Abnormal noise from the mechanism
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Trembles during movement
Dirt in linear bearings or driving
surface.
Spiral gear damaged.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Abnormal noise from the mech-
anism
Cassette carriage touches sec-
ondary slot or other construc-
tions.
Inner secondary slot touches
the outer secondary slot in con-
nection with the step motor res-
onance.
Cassette glide/adhesive strips
resonate.