Chapter D - ADJUSTMENT & CALIBRATION
D-4 Planmeca Proline EC X-ray unit
PANORAMIC X-RAY BEAM ADJUSTMENT
Technical Manual
2 PANORAMIC X-RAY BEAM ADJUSTMENT
NOTE
In case the X-ray unit is equipped with optional AEC (Automatic Exposure Control), the
AEC calibration must be performed after the panoramic X-ray beam adjustment. See
instructions given in section “AUTOMATIC EXPOSURE CONTROL (AEC) CALIBRA-
TION” on page D-27.
2.1 Checking the panoramic beam position
NOTE
These instructions are valid for both the Planmeca Panoramic EC X-ray and Planmeca
Panoramic EC Pan/Ceph X-ray.
Before the X-ray is used the position of the radiation beam must be checked. Remove the
inner cover from tubehead assembly. Four screws hold the tubehead cover in position, these
are located at the rear of the tubehead assembly. The primary slot mechanism is located
under the tubehead cover.
Figure 1
Switch the unit on. Enter the test mode (see section “Test mode” on page B-4).
Press the ready key to drive the X-ray into the exposure position. The rotating unit and cas-
sette carriage will move to their respective ready positions. When the X-ray is ready the indi-
cator light will come on.
Start to take a test exposure by pressing and holding down the exposure button on the
remote control. Stop pressing the exposure button when the tubehead has rotated to a con-
venient position for viewing the secondary slot.
An error code will appear on the display and start to flash.
Press the CTL-key to clear the error code.
Tubehead cover
Cassette drive assembly
READY
PRET
Er 00
CTL
Summary of Contents for Proline EC ceph
Page 2: ......
Page 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Page 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Page 208: ......
Page 209: ......