Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline EC X-ray unit E-69
IMAGE AREA SHIFTING
Technical Manual
The parameter value CS1 indicates the “old” image area (moves the image area backwards).
The parameter value CS2 moves the image area forwards approx. 14mm (0.55 in.) regarding
to the image area with value CS1. Change the parameter value with the up/down keys.
To accept the parameter value press the Ready key.
To exit the service mode press the CEPH key briefly.
You have to adjust the cephalometric beam after changing the parameter value. Refer to the
panoramic unit’s technical manual.
NOTE
If you do not want to move the image area when upgrading the software from the revi-
sions smaller than 5.12, select the parameter value CS=1. In that case you do not have
to adjust the cephalometric beam.
READY