Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline EC X-ray unit D-17
PATIENT POSITIONING MECHANISM
Technical Manual
Leave the ball phantom tool in position and place the alignment ruler in position between the
primary slot and the secondary slot as described earlier. The line on the alignment rule must
not line up with the line on the ball phantom but must be parallel to it. (If the two lines do line
up it indicates that the patient positioning mechanism is correctly positioned).
Figure 24
Enter the service mode.
Press either of the patient positioning keys once to switch the patient positioning lights on
then a second time to adjust the position of the patient positioning mechanism until the line on
the ball phantom lines up with the line on the alignment ruler. The position of the mechanism
will appear on the display.
Press the CTL-key. The indicator light will come on.
Hold the patient positioning plus (+) key down (about five seconds) until you hear a signal and
the millimeter display zeroes. This indicates that the new position of the patient positioning
mechanism has been programmed into the memory.
Remove the alignment pin and exit the service mode.
Take another ball phantom picture to check the alignment again. Readjust if necessary.
Alignment ruler
Ball phantom
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
X-LINECA.EPS
mm
12
CTL
mm
00
Summary of Contents for Proline EC ceph
Page 2: ......
Page 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Page 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Page 208: ......
Page 209: ......