Chapter G - TROUBLESHOOTING
Planmeca Proline EC X-ray unit G-11
PROBLEMS IN RADIOGRAPHS
Technical Manual
9.3 Exposure on - no picture
9.4 Very light radiographs
9.5 Stripes on the film
9.6 Exposed area incorrectly limited
Film has certain tolerance in the holder which causes considerable variations.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Film unexposed
Primary slot does not let in radi-
ation to film.
In the developer the fixative
and developer changed places.
Film completely black
Light leakages to dark room.
Film otherwise exposed.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Very light radiographs
Film cassette upside down in
its holder.
Primary slot positioned so that
the radiation beam goes trough
the secondary slot.
Secondary slot moved.
Wrong film/intensifying film
combination.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Horizontal stripes
From developing machine.
Primary slot tilted so that part of
the film meets secondary slot.
Patient has jewels, large fillings
in teeth, etc.
Vertical stripes
From rolls of developing
machine.
Cassette mechanism gets tuck.
Generator working improperly.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Radiograph incorrectly limited
in vertical direction.
Child limit chosen by accident.
Primary slot limits adjusted
incorrectly.
Summary of Contents for Proline EC ceph
Page 2: ......
Page 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Page 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Page 208: ......
Page 209: ......