Maintenance, Cleaning, Troubleshooting, and Disposal
Maintenance
Maintenance run
If the platform only moves in a limited travel range, we recommend running once over the
entire range with the following intervals (maintenance run). This allows the lubricant to be
distributed evenly over the relevant machine elements.
Type of operation
Interval
Laboratory use
After 50 operating hours or at least after one 1 year
Continuous industrial
operation
After every 5000 motion cycles
Relubricating
Type of operation
Interval
Laboratory use
Normally not necessary
Continuous industrial
operation
According to separately agreed maintenance schedule
If you have any questions on relubricating, contact our customer service department
Cleaning
Tools
Soft cloth
Mild cleaning agent or disinfectant
Procedure
INFORMATION
Do not allow the cleaning agent or disinfectant to get onto the drive screw or guides. This
could impair lubrication.
1.
Remove the motor cable from the controller.
2.
Dampen the cloth with the cleaning agent or disinfectant.
3.
Wipe over the dirty surface gently.
4.
After cleaning, reconnect the motor cable to the controller and tighten the screws to
secure the plug.