50 Specifications
Philips Respironics E30 User Manual
7.1.6 Electrical
AC Power Consumption (with 80W power supply)
Note: Power supply is part of the medical electrical
equipment.
100 – 240 VAC, 50/60 Hz, 2.0-1.0 A
DC Power Consumption
12 VDC, 6.67 A
Fuses
There are no user-replaceable fuses.
7.1.7 Intake port filters
Pollen Filter
100% Polyester
88% Efficient @ 7-10 micron size
Ultra-fine Filter
Blended Synthetic Fiber
95% Efficient @ 0.5-0.7 micron size
7.1.8 Control accuracy
Parameter
Range
Accuracy
IPAP
4 - 30 cmH
2
O
± 2.5 cmH
2
O*
EPAP
4 - 25 cmH
2
O
± 2.5 cmH
2
O*
CPAP
4 - 20 cmH
2
O
± 2.5 cmH
2
O*
Breath rate
0 to 30 BPM
greater of ± 1 BPM or ±10% of setting
Inspiration time
0.5 to 3 seconds
± (10% of s 0.1 second)
*Pressure measured at the patient connection port with or without the integrated heated humidifer (no patient flow).
7.1.9 Pressure accuracy
Pressure increments: 4.0 to 30.0 cmH
2
O (in 0.5 cmH
2
O increments)
Maximum static pressure accuracy, according to ISO 80601-2-70:2015:
Parameter
Static Accuracy
10 cmH
2
O*
± 3.0 cmH
2
O
Static pressure accuracy has a measurement uncertainty of 3.7%
Maximum dynamic pressure variation, according to ISO 80601-2-70:2015:
Pressure
10 BPM
15 BPM
20 BPM
< 10 cmH
2
O
± 0.4 cmH
2
O
± 0.5 cmH
2
O
± 0.8 cmH
2
O
≥
10.0 to 20 cmH
2
O
± 0.5 cmH
2
O
± 0.8 cmH
2
O
± 1.0 cmH
2
O
Dynamic pressure accuracy has a measurement uncertainty of 4.3%
Summary of Contents for 1146562
Page 2: ... 2020 Koninklijke Philips N V All rights reserved ...
Page 18: ...12 Introduction Philips Respironics E30 User Manual ...
Page 20: ...14 Therapy modes Philips Respironics E30 User Manual ...
Page 54: ...48 Cleaning Philips Respironics E30 User Manual ...
Page 60: ...54 Specifications Philips Respironics E30 User Manual ...
Page 63: ...Limited warranty 57 Philips Respironics E30 User Manual ...
Page 64: ...58 Limited warranty Philips Respironics E30 User Manual ...