Alarms, alerts, and troubleshooting 35
Philips Respironics E30 User Manual
Alert
Icon
Priority
Description
User action
Oximetry: Good
Connection
(icon only)
Status
Displays on the therapy
screen when the blower is
on and 3 seconds of good
connection is detected.
Appears at the beginning
of therapy. This screen will
not display again if the
oximetry sensor is removed
and reapplied unless
therapy is stopped and
restarted.
No action needed.
Flow Sensor Error:
Limited Therapy
Notification
Flow sensor malfunction.
Press the control dial to
acknowledge and clear
the message. Disconnect
the device from power.
Immediately remove patient
from ventilator and connect
them to alternate source of
ventilation. Contact Philips
Respironics or the product
support team.
SD Card Removed.
Notification
or Alert
Indicates SD card has been
removed from therapy
device and not reinserted
before the start of the
current therapy session.
Reinsert SD card, or press the
control dial to clear alert.
Oximetry: Good
Study
(icon only)
Notification
After exiting therapy, the
therapy device will display
the Good Oximetry pop-
up screen if the oximetry
sensor maintained a good
connection for at least 4
consecutive hours within a
24 hour window with breaks
in therapy no greater than 1
hour each.
Press the control dial to
acknowledge and clear the
message.
Summary of Contents for 1146562
Page 2: ... 2020 Koninklijke Philips N V All rights reserved ...
Page 18: ...12 Introduction Philips Respironics E30 User Manual ...
Page 20: ...14 Therapy modes Philips Respironics E30 User Manual ...
Page 54: ...48 Cleaning Philips Respironics E30 User Manual ...
Page 60: ...54 Specifications Philips Respironics E30 User Manual ...
Page 63: ...Limited warranty 57 Philips Respironics E30 User Manual ...
Page 64: ...58 Limited warranty Philips Respironics E30 User Manual ...