Cleaning 47
Philips Respironics E30 User Manual
4. Use one of the following to disinfect all exterior surfaces of the device and O
2
inlet port:
—
70% Isopropyl Alcohol (Device and O
2
Inlet Port)
—
DisCide Towelettes (Device ONLY)
—
10% Chlorine Bleach solution (Device ONLY)
5. Pay close attention to all corners and crevices.
6. Allow the device to dry completely before plugging in the power cord and re-inserting the filter(s).
6.4 Service
The device does not require routine servicing or preventative maintenance other than inspection for normal
wear or damage and device cleaning.
Warning
If you notice any unexplained changes in the performance of this device, if it is making unusual
or harsh sounds, if it has been dropped or mishandled, if water is spilled into the enc
losure,
or if the enclosure is broken, disconnect the power cord and discontinue use. Contact Philips
Respironics.