36
Alarms, alerts, and troubleshooting
Philips Respironics E30 User Manual
Alert
Icon
Priority
Description
User action
SD Card Error:
Remove and
Reinsert
Notification
SD card error detected.
Device cannot read the SD
card.
A problem may exist with
the SD card or it was
ejected during a writing
activity, or it was inserted
incorrectly.
Remove SD card and reinsert.
If alert continues to occur,
replace with another card or
contact Philips Respironics or
the product support team.
SD Card Full.
Notification
SD card is full.
Remove SD card and contact
Philips Respironics or the
product support team for a
new SD card.
Humidification
Error. Contact
support if the
problem persists.
Status
Humidifier error (only when
humidifier is present)
Humidifier heater plate
error or humidifier not
properly connected to
therapy device.
Turn off device and
disconnect from power.
Detach the humidifier,
visually check that electrical
contacts are clear, then
reconnect humidifier
and power cord. If alert
continues, contact Philips
Respironics or the product
support team.
The attached
power supply
does not support
humidification.
Alert
Indicates that the attached
power supply is not
capable of supporting
humidification.
Contact Philips Respironics
or the product support team
to obtain a compatible
Philips Respironics power
supply. Or, operate therapy
device without humidifier.
Activate MODE?
Notification
Indicates mode change
while blower is running or
therapy is being delivered.
Choose Yes to accept new
therapy mode, No to stay in
current mode.
Check Power
Notification
Indicates an incompatible
power supply is attached.
Incompatible power supply,
or power cord is not fully
inserted into device’s power
inlet.
Confirm power cord is fully
inserted into device’s power
inlet. Confirm a compatible
Philips Respironics power
supply is attached. Switch to
compatible power supply if
needed.
Busy
Status
Displayed when the device
is temporarily inaccessible
due to data communication.
No action needed.
Summary of Contents for 1146562
Page 2: ... 2020 Koninklijke Philips N V All rights reserved ...
Page 18: ...12 Introduction Philips Respironics E30 User Manual ...
Page 20: ...14 Therapy modes Philips Respironics E30 User Manual ...
Page 54: ...48 Cleaning Philips Respironics E30 User Manual ...
Page 60: ...54 Specifications Philips Respironics E30 User Manual ...
Page 63: ...Limited warranty 57 Philips Respironics E30 User Manual ...
Page 64: ...58 Limited warranty Philips Respironics E30 User Manual ...