Precision 500
User Manual
Page 25 2012 © Peak Scientific – Rev 2 – 07/11/12
Expanding Capacity
The Precision 500 has an expandable feature, where the flow capacity of the
unit is increased, by connecting up to 16 units in a Master-Slave format.
Rear Connections
In order to use this feature the units must be connected together via the
‘Master’ port on the first Precision 500, and the ‘Slave’ port at the rear of the
next Precision 500 unit.
The first Precision 500 unit is automatically defined as the Master, and before
the connections are made and H
2
is supplied, the desired pressure must be set
on all units. This must be the same pressure for all.
The user can now press Start on the Master unit to begin the flow of H
2
. The
Master will now manage all the other units.
The Data Cables (08-8902) used to connect units together can be purchased
from Peak Scientific.
Master Port
Slave Port
Slave
EOL
Master