Installation
Location
The generator should be located on a flat, level surface for operation - this will
ensure that the water level sensors within the unit operate correctly. Some
components in the generator produce heat during operation and therefore an
adequate airflow must be maintained around the generator to allow the
ventilation system to perform efficiently. The rear of the generator will become
warm to the touch during operation - a minimum clearance of 15 cm (6") from
other bodies is recommended. Do not locate the generator in a sealed or
unventilated room, or in close proximity to any ignition sources.
Caution - Remove Transport Plugs
Before installation, remove the blanking plugs fitted to the vents at the rear of
the generator. Do not obstruct or connect the vents to any application. Failing
to observe this could lead to permanent damage of the generator.
Please ensure that the Transport plugs are retained for future transportation.
Do not expose the unit to freezing conditions. Please adhere to the operating
temperatures below:
Minimum Operating Ambient Temperature:
10 °C (50 °F)
Maximum Operating Ambient Temperature:
25 °C (77 °F)
Note: It is recommended that the unit is left to run for 12 hours before
utilisation.