96
4041
OverServ
w/
r
0...1
0
Analog output overflow management
Value
Description
0
Switched off
1
Switched on
4042
Mode
w/
r
0...2
0
RS-485 interface working mode
0
RS-485 interface works in modbus
Slave mode, transducer waits for requ-
ests and responds on request adressed
to its address
1
Transducer monitors trafic on RS-485
interface and acts on data other devices
transmitt
2
RS-485 interface works in modbus
Master mode, transducer sends requests
and waits for reply from Slave device
4043
Address
w/
r
0...247
1
RS-485
MODBUS
network
address.
Enter 0 to switch off the interface.
4044
ModeUnit
w/
r
0...3
0
RS-485 interface transmission mode
0
RTU 8N2
1
RTU 8E1
2
RTU 8O1
3
RTU 8N1
4045
BaudRate
w/
r
0...7
1
RS-485 interface transmission baudrate
Value
Description
0
4800 bit/s
1
9600 bit/s
2
19200 bit/s
3
38400 bit/s
4
57600 bit/s
5
115200 bit/s
6
230400 bit/s
7
256000 bit/s