86
Device address
Function
Number of bytes
Device ID
Device state
Device-dependent field
CRC
Firm-
ware
v 2.00
Registers 4304, 4305
describing the serial
number and hardware
configuration of the
transducer
(serial no 13100001)
01h
11h
08h
C1h
FFh
02h 00h
A0h 01h 6Ch 0Dh
69FCh
Table 35
Response:
Device-dependent field – 4 bytes corresponding to register value
4304, 4305 see table 41 manufacturing status 1,
manufacturing status 2.
5.9.4. Register map
In the PCE-P30U transducer the data is stored in 16- and 32-bit
registers. The process variables and parameters of the device are
stored in the different address space depending on the variable type.
The bits in the 16-bit registers are numbered from the least significant
to the most significant (b0 … b15). The 32-bit registers (4 Bytes) contain
floating-point values in IEEE-754 standard. Bytes sequence: B3 B2 B1
B0 – the most significant byte is sent as the first one. 16-bit registers
which represents 32-bit values on a two following registers are multiplied
at different address field with different bytes (word) order: B1 B0 B3 B2
(table. 36).
Register map of the PCE-P30U transducer is shown in table 36.
Note:
All the given addresses are physical addresses. In some
computer programs logical addressing is applied, then the addresses
should be increased by 1.