72
Table 20
Register
HEX value
4553
0x0170
4554
0xBB95
4555
0xE87C
4556
0xB942
rec = 0x
0170BB95
E87CB942
Data = 0xE87CB942 → (float) → 92.743958;
Table 21
Recording time =
0x0170BB95
→ b1011100001011101110010101
Year + 2010
Month
Day
Hour
Minute
Second
6 bytes
4 bytes
5 bytes
5 bytes
6 bytes
6 bytes
0 0 0 0 0 0 0
1
0
1 1 1
0 0 0 0
1
0
1 1 1
0
1 1 1
0 0
1
0
1
0
1
0 + 2010
5
24
11
46
21
10-05-24 11:46
Rec : 2010-05-24 11:46:21 92.743958
5.8.2.2. Downloading archived data from the internal memory
Downloading of archive data from the internal memory is performed via
the memory card (option) or via the RS-485 interface. Downloading data
consists in reading subsequent memory pages containing data records.
eCon software enables acquiring individual pages from the internal me-
mory.
If the transducer has been manufactured in a variant supporting
external SD/SDHC cards, then the archive data can automatically
be copied to the memory card (this is the fastest method of obtaining