110
5.9.6. Read-only registers
Table 42
Register address
(16 bit registers)
W
rite (w)/ read (r)
Range
Description
4300
r
0...9999
Firmware version * 100
4301
r
0...65535
Transducer status 1. Describes the current transdu
-
cer status. The consecutive bits represent a given
event. Bit set to 1 means that the event has taken
place. Events can only be cleared.
Bit15
31
Loss of calibration parameters
Bit14
30
RTC – loss of presets – battery
error
Bit13
29
Clock – change of winter/sum
-
mer time
Bit12
28
No communication with data
memory
Bit11
27
Wrong settings
Bit10
26
Default settings have been
restored
Bit9
25
Measurement range overflow
Bit8
24
Error in communication with
internal archive memory
Bit7
23
Archive parameters error
Bit6
22
ADC converter error
Bit5
21
100% use of the internal memory
archive