20
6
Method of
reactive
energy
calculation
en_q
cAP, sIgn
cAP – inductive and capacity
energy
sIgn – positive and negative
energy
cAP
7
Display
panel illumi-
nation
diSP
oFF,1…60,
on
off, on, 1..60 – illumination
time (in seconds) from pres-
sing the button
on
8
Energy
counters
erasing
En_0 no, EnP, Enq,
ALL
no – no activity,
EnP – erase active energy,
Enq – erase reactive,
ALL – erase all energies
no
9
Mean acti-
ve power
erasing
PA_0
no, yES
yES – erase power
no
10
Ordered
power
PAor
0 … 144.0
Power ordered to establish
power consumption in % of
rated power
100
11
Default
settings
dEf
no, yES
Reverting to default (factory)
group settings
no
The automatic erasing of energy is carried out:
- for active energy when changing: voltage or current ratio;
- for reactive energy when changing: voltage or current ratio,
reactive energy calculation method;
Buttons
and
are used for setting the values while
buttons
and
are used for choosing position
of the number to be set. The active position is signaled by the
cursor. Set value can be accepted by pressing the
button
or canceled by pressing the
button. When value is to
be accepted, it is checked against the acceptable value range.
If the set value falls outside the allowable range, the meter
remains in parameter setting mode and the value is set to
the highest possible value (when entered value is too high)
or lowest possible value (when it is too low).
Summary of Contents for PCE-ND10
Page 7: ...9 5 3 Connection diagrams OUT Direct measurement in a 4 wire network PCE ND10...
Page 8: ...10 OUTimp Semi indirect measurement in a 4 wire network PCE ND10...
Page 15: ...17 6 4 Operating modes Fig 6 PCE ND10 meter operating modes...
Page 24: ...26 Fig 8 Alarm types x alarm no a b normal c off d on a n on b n oFF c On d OFF...