Chapter 1 – Installation
12 Wait for the first internal calibration to complete.
The display indicates that the weight on the pan is less than the expected weight of at least
200 grams.
13 Place a weight between 200 and 600 grams on the center of the weighing pan.
The scale continues with external calibration.
14 Wait for the external calibration to complete, after which the scale flashes
> ----
and displays
...Measuring zero...
because the weight on the pan is now greater than the expected weight of
0 grams.
15 Remove the weight from the weighing pan and wait for the scale to perform one more internal
calibration.
The scale displays the
Scale needs to be sealed?
prompt.