background image

3

2

 INTENDED USE/SPECIFICATIONS

Intended use

z

„

This stretcher is designed to be used in medical institutions to transport patients or support 
them as they receive treatment.

*  

For details of the patient target groups, the intended users and the clinical benefits, see “Specifications” below.

Specifications

Description

Transfer Stretcher

Model No.

KK-728EG

Category/Classification

Mechanical implement 1: Operating table and treatment table,  

general diagnostic and processing table 13958000

Ordinary medical devices and specially maintained and controlled medical devices

Dimensions

Total length

193cm

Total width

66.3cm

Mattress base height (cm) (*1)

51 - 85cm

Back section base Inclination angle

0º to 70°

Casters

150mm dia. Casters (with central locking system) with center wheel

Total weight

71kg (not including the transfer mattress)

maximum patient weight

135kg

Safe working load (*2)

1500N (153kgf)

Service life

8 years (based on Paramount data), excluding consumable items such as casters.

Mattress

Dimensions

190cm × 53cm × 2cm

Surface material

Polyester

Padding

Urethane foam

Materials

Frame

Steel/partially aluminum

Base

Polypropylene (PP)

Transfer board

Polypropylene (PP)

Capacity of compatible Oxygen cylinder

500 L (diameter 10cm to 11.5cm)

Accessories

IV Pole (C57E)  Load capacity: 1kg/hook

Oxygen cylinder holder  Load capacity: 15kg

Tray: Polypropylene (PP)  Load capacity: 10kg

Hook (3K003355: 4 pieces)  Load capacity: 3kg/piece

Patient target groups

Patient age

13 years of age or older

Maximum patient weight

135kg

Intended users

Health care workers, caregivers, and maintenance staff

*1)  The mattress bas

e height indicates the height from the floor to the mattress base surface.

*2)  Safe working load: Maximum weight that the stretcher can safely accept (the total load of the patient’s weight and 

incidental items including the mattress and other accessories)

Environmental conditions

Range

„

 During transportation or storage

„

 During operating

Temperature

-10ºC to 50ºC

0ºC to 40ºC

Humidity

30%RH to 90%RH

30%RH to 90%RH

Atmospheric pressure

700hPa to 1060hPa

700hPa to 1060hPa

·  Essential performance

Function to support and transfer the body of patients

·  Frequently used functions 

Caster lock/unlock, transferring, change of transfer board position.

·

  Clinical benefits 

 

Easy patient transport in hospitals  

Safe and efficient patient transfer to hospital beds 

 

Easy and safe treatment reception of the patient

·  Applicable standards

The stretcher described in this instruction manual is designed and manufactured according to the following equipment 

classification and standards.

Applicable standards:  IEC60601-1:2005 + A1:2012 

EN60601-1:2006 + A12:2014

Classification as medical device in Europe

: Class I

Summary of Contents for KK-728EG

Page 1: ...Serious incident means any incident that directly or indirectly led might have led or might lead to any of the following a the death of a patient user or other person b the temporary or permanent ser...

Page 2: ...wheel system 11 3 Height adjustment 12 4 Raising the backrest 13 5 Transfer board 14 6 Restraint belt 16 7 Oxygen cylinder holder 17 8 IV Pole C57E 19 9 Accessory rail 21 10 Transfer mattress 21 6 PAT...

Page 3: ...ction base Accessory rail Oxygen cylinder holder Caster lock pedals Tray Base frame Height adjusting handle Main frame Leg section base IV Pole attachment hole Transfer board Restraint belt Transfer m...

Page 4: ...es IV Pole C57E Load capacity 1kg hook Oxygen cylinder holder Load capacity 15kg Tray Polypropylene PP Load capacity 10kg Hook 3K003355 4 pieces Load capacity 3kg piece Patient target groups Patient a...

Page 5: ...n could cause death or serious injury broken bones pressure or paralysis CAUTION Indicates that improper handling by ignoring this indication could cause light personal injury bruises wounds or cuts o...

Page 6: ...on the stretcher z z There is a risk of injury caused by falling or toppling from the stretcher In particular be careful with children Do not crawl under or place your head hands or legs under the st...

Page 7: ...rely secure the patient if using it with the two transfer boards locked in the horizontal level position z z Otherwise the patient could lose balance and fall out of the stretcher or fall over resulti...

Page 8: ...e stretcher Otherwise the stretcher may be damaged or deformed or even catch fire Use care when handling the gas spring z z Gas spring is used for the backrest lever If you attempt to disassemble or i...

Page 9: ...her is not in use store it with the casters facing in z z Otherwise people may trip over a caster and fall down or get injured Do not place objects weighing 10kg or more on the tray z z The tray could...

Page 10: ...ve amounts of water vapor or oil vapor z z Locations that tend to be exposed to high temperatures or humidity low temperatures or dry conditions Note The optimum ambient conditions for the stretcher a...

Page 11: ...operating only one caster lock pedal Caster lock pedals are located at the four corners of the stretcher allowing them to be operated at each side of head and foot ends To lock the casters z z Depress...

Page 12: ...stretcher may tip over resulting in an injury When the center wheel is OFF z z Depressing the center wheel locking pedal sets the center wheel to OFF to allow the stretcher to move in any direction in...

Page 13: ...heavy when it is turned close to the limit Do not turn the handle over the operating range Turning the handle forcibly over the range may result in damage or malfunction z z Do not put a leg on the ha...

Page 14: ...equipment Grip Grip Do not hold any other place besides the grip To raise z z Hold the grip of the back section base with one hand hold the backrest lever on the back section base with your other han...

Page 15: ...ay fall off and get injured To lower the transfer board To raise the transfer board To raise the transfer board 1 Slowly lift the transfer board which has been lowered in the direction shown in the fi...

Page 16: ...wly lift the transfer board to a horizontal level position See To raise the transfer board in p 14 2 Move the transfer board in the direction shown by the arrow while keeping it in a horizontal level...

Page 17: ...ay fall off or the stretcher may tip over resulting in an injury z z After placing a patient on the base back section base and leg section base of the stretcher use the two pairs of restraint belts to...

Page 18: ...nd flowmeter may be deformed resulting in injury 7 Oxygen cylinder holder The oxygen cylinder holder can contain a 500L 10cm to 11 5cm diameter oxygen cylinder The oxygen cylinder holder can be set ei...

Page 19: ...therwise the oxygen cylinder holder and flowmeter may be deformed resulting in injury Rotation stopper fitting Stopper Attachment part for horizontal setting Boss Thumb screw knob for locking a cylind...

Page 20: ...g the sliding pipe of the IV Pole adjust the sliding pipe length and then lock the sliding pipe by tightening the loosened thumb screw knob WARNING z z When loosening the thumb screw knob firmly grip...

Page 21: ...sliding pipe is loosened the sliding pipe may unintentionally fall resulting in injury z z Before use make sure the thumb screw knob for locking the sliding pipe is firmly tightened The sliding pipe m...

Page 22: ...the hook or accessory rail resulting in injury 10 Transfer mattress A patient can be transferred to from a bed without being lifted For the procedures see 6 PATIENT TRANSFER on p 22 to 25 The mattress...

Page 23: ...y or the side rail of the bed or the transfer board may get damaged or deformed Transfer board Transfer board Transfer board Transfer board receptacle When the transfer board is up When the transfer b...

Page 24: ...their hands together on their abdomen and pull out the side part of the transfer mattress that was folded in step 4 Grab handles of the transfer mattress handles on the stretcher side and pull the mat...

Page 25: ...ck pedal green to unlock the casters move the stretcher slightly away from the bed and raise the transfer board see p 14 WARNING z z When operating the transfer board place the patient s hands togethe...

Page 26: ...they will not hinder the transfer procedure Grab the handles of the transfer mattress and transfer the patient onto the bed by slowly pulling the mattress 4 After transferring the patient 1 Fold the s...

Page 27: ...as a support to stand up z z Do not transport a patient with the transfer board locked in the horizontal level position z z When transporting a patient be sure to fasten the restraint belts z z Operat...

Page 28: ...bnormal noise If water gets on them immediately wipe it off Maintenance and inspection items to be completed by the user of the product Main Unit The position operations using the height adjusting han...

Page 29: ...nium chloride e g Osvan Up to 0 2 Benzethonium chloride e g Hiamine Up to 0 05 Chlorhexidine gluconate e g Hibitane Up to 0 05 Sodium hypochlorite e g Milton It is highly recommended that you thorough...

Page 30: ...e center wheel is set to ON z z Turn OFF the center wheel 11 The transfer board cannot be lowered z z Check if the safety lock is ON z z Release the safety lock 14 15 The transfer board rattles z z Ch...

Page 31: ...od Paramount Bed keeps replacement parts necessary to maintain stretcher performance in stock for at least 8 years after the discontinuation of production 3 Questions regarding after sales service Con...

Page 32: ...Translation of the original Instruction Made by Torindo Co Ltd 2022 04...