To store

1

 Close the hooks and adjust the sliding pipe to the 

shortest length.

2

 Align the thumb screw knob (for locking the sliding 

pipe) of the IV Pole with the groove on the IV Pole 
storage holder or tray and store away the IV Pole.

2

 Open the hooks at the tip when using the IV Pole.

NOTE!

z

z

The load capacity of each hook is 1kg.

Hooks

 WARNING

z

z

When using the IV Pole, do not put a hand or arm between the base or transfer 
board and the IV Pole. The body part may get caught, resulting in injury.

z

z

The load capacity of each hook is 1kg. Do not hang an article exceeding the 
load capacity because it may damage the hook, resulting in injury. After inserting 
an oxygen cylinder into the oxygen cylinder holder, tighten the thumb screw 
knobs properly. Otherwise, the oxygen cylinder may move unexpectedly or drop, 
resulting in an injury or damage.

z

z

Do not adjust the sliding pipe length while articles are hanging on the hooks. 
When the thumb screw knob (for locking the sliding pipe) is loosened, the sliding 
pipe may unintentionally fall, resulting in injury.

z

z

Before use, make sure the thumb screw knob (for locking the sliding pipe) is firmly 

tightened. The sliding pipe may unintentionally fall, resulting in injury.

z

z

Attach an infusion bag or other article to a hook properly. The infusion bag or other 
article may unexpectedly fall, resulting in injury or damage to the infusion bag or 
other article.

 WARNING

z

z

When not using an IV Pole, close the hooks, adjust the IV Pole to the shortest 
length, and store the IV Pole in the IV Pole storage holder in the lower part of 
the back section base or in the IV Pole storage space on the tray. If not stored 
properly, the IV Pole could hit other parts when the height of the stretcher is 
adjusted, causing damage or deformation of the IV Pole or stretcher.

Thumb screw knob 
(for locking an IV Pole)

Back section base

Align the thumb screw knob (for locking the sliding pipe) 
with the groove on the holder and store away the IV Pole.

Hooks (folded when stored)

IV Pole (C57E)

IV Pole storage holder
* Can be stored here when it is not in use.

IV Pole attachment hole

Hooks

*  An IV Pole can be stored 

in the tray.

Summary of Contents for KK-728EG

Page 1: ...Serious incident means any incident that directly or indirectly led might have led or might lead to any of the following a the death of a patient user or other person b the temporary or permanent ser...

Page 2: ...wheel system 11 3 Height adjustment 12 4 Raising the backrest 13 5 Transfer board 14 6 Restraint belt 16 7 Oxygen cylinder holder 17 8 IV Pole C57E 19 9 Accessory rail 21 10 Transfer mattress 21 6 PAT...

Page 3: ...ction base Accessory rail Oxygen cylinder holder Caster lock pedals Tray Base frame Height adjusting handle Main frame Leg section base IV Pole attachment hole Transfer board Restraint belt Transfer m...

Page 4: ...es IV Pole C57E Load capacity 1kg hook Oxygen cylinder holder Load capacity 15kg Tray Polypropylene PP Load capacity 10kg Hook 3K003355 4 pieces Load capacity 3kg piece Patient target groups Patient a...

Page 5: ...n could cause death or serious injury broken bones pressure or paralysis CAUTION Indicates that improper handling by ignoring this indication could cause light personal injury bruises wounds or cuts o...

Page 6: ...on the stretcher z z There is a risk of injury caused by falling or toppling from the stretcher In particular be careful with children Do not crawl under or place your head hands or legs under the st...

Page 7: ...rely secure the patient if using it with the two transfer boards locked in the horizontal level position z z Otherwise the patient could lose balance and fall out of the stretcher or fall over resulti...

Page 8: ...e stretcher Otherwise the stretcher may be damaged or deformed or even catch fire Use care when handling the gas spring z z Gas spring is used for the backrest lever If you attempt to disassemble or i...

Page 9: ...her is not in use store it with the casters facing in z z Otherwise people may trip over a caster and fall down or get injured Do not place objects weighing 10kg or more on the tray z z The tray could...

Page 10: ...ve amounts of water vapor or oil vapor z z Locations that tend to be exposed to high temperatures or humidity low temperatures or dry conditions Note The optimum ambient conditions for the stretcher a...

Page 11: ...operating only one caster lock pedal Caster lock pedals are located at the four corners of the stretcher allowing them to be operated at each side of head and foot ends To lock the casters z z Depress...

Page 12: ...stretcher may tip over resulting in an injury When the center wheel is OFF z z Depressing the center wheel locking pedal sets the center wheel to OFF to allow the stretcher to move in any direction in...

Page 13: ...heavy when it is turned close to the limit Do not turn the handle over the operating range Turning the handle forcibly over the range may result in damage or malfunction z z Do not put a leg on the ha...

Page 14: ...equipment Grip Grip Do not hold any other place besides the grip To raise z z Hold the grip of the back section base with one hand hold the backrest lever on the back section base with your other han...

Page 15: ...ay fall off and get injured To lower the transfer board To raise the transfer board To raise the transfer board 1 Slowly lift the transfer board which has been lowered in the direction shown in the fi...

Page 16: ...wly lift the transfer board to a horizontal level position See To raise the transfer board in p 14 2 Move the transfer board in the direction shown by the arrow while keeping it in a horizontal level...

Page 17: ...ay fall off or the stretcher may tip over resulting in an injury z z After placing a patient on the base back section base and leg section base of the stretcher use the two pairs of restraint belts to...

Page 18: ...nd flowmeter may be deformed resulting in injury 7 Oxygen cylinder holder The oxygen cylinder holder can contain a 500L 10cm to 11 5cm diameter oxygen cylinder The oxygen cylinder holder can be set ei...

Page 19: ...therwise the oxygen cylinder holder and flowmeter may be deformed resulting in injury Rotation stopper fitting Stopper Attachment part for horizontal setting Boss Thumb screw knob for locking a cylind...

Page 20: ...g the sliding pipe of the IV Pole adjust the sliding pipe length and then lock the sliding pipe by tightening the loosened thumb screw knob WARNING z z When loosening the thumb screw knob firmly grip...

Page 21: ...sliding pipe is loosened the sliding pipe may unintentionally fall resulting in injury z z Before use make sure the thumb screw knob for locking the sliding pipe is firmly tightened The sliding pipe m...

Page 22: ...the hook or accessory rail resulting in injury 10 Transfer mattress A patient can be transferred to from a bed without being lifted For the procedures see 6 PATIENT TRANSFER on p 22 to 25 The mattress...

Page 23: ...y or the side rail of the bed or the transfer board may get damaged or deformed Transfer board Transfer board Transfer board Transfer board receptacle When the transfer board is up When the transfer b...

Page 24: ...their hands together on their abdomen and pull out the side part of the transfer mattress that was folded in step 4 Grab handles of the transfer mattress handles on the stretcher side and pull the mat...

Page 25: ...ck pedal green to unlock the casters move the stretcher slightly away from the bed and raise the transfer board see p 14 WARNING z z When operating the transfer board place the patient s hands togethe...

Page 26: ...they will not hinder the transfer procedure Grab the handles of the transfer mattress and transfer the patient onto the bed by slowly pulling the mattress 4 After transferring the patient 1 Fold the s...

Page 27: ...as a support to stand up z z Do not transport a patient with the transfer board locked in the horizontal level position z z When transporting a patient be sure to fasten the restraint belts z z Operat...

Page 28: ...bnormal noise If water gets on them immediately wipe it off Maintenance and inspection items to be completed by the user of the product Main Unit The position operations using the height adjusting han...

Page 29: ...nium chloride e g Osvan Up to 0 2 Benzethonium chloride e g Hiamine Up to 0 05 Chlorhexidine gluconate e g Hibitane Up to 0 05 Sodium hypochlorite e g Milton It is highly recommended that you thorough...

Page 30: ...e center wheel is set to ON z z Turn OFF the center wheel 11 The transfer board cannot be lowered z z Check if the safety lock is ON z z Release the safety lock 14 15 The transfer board rattles z z Ch...

Page 31: ...od Paramount Bed keeps replacement parts necessary to maintain stretcher performance in stock for at least 8 years after the discontinuation of production 3 Questions regarding after sales service Con...

Page 32: ...Translation of the original Instruction Made by Torindo Co Ltd 2022 04...

Reviews: