PARAMOUNT BED KK-728EG Instruction Manual Download Page 15

Foot end

Head end

Foot end

Head end

14

5. Transfer board

z

„

Transfer boards can be used to protect a patient on the stretcher from falling off and also 

as a bridge between the stretcher and a bed when transferring a patient to/from a bed. 
They can be used as an arm table by locking them in the horizontal level position.

 WARNING

z

z

When patients are placed on the base of stretcher, secure 
them with the restraint belts (see p. 16), raise the transfer 
boards, and continuously monitor them. 

Otherwise, the patient may fall off from the 

stretcher and get injured.

z

z

After raising the transfer boards or locking 
them in the horizontal level position, use 

the safety lock and confirm that they are 

properly locked (Green in the indicator 
window). If they are not locked properly (red 
shows in the indicator window), the transfer boards may lower 

and the patient may fall off and get injured.

z

„

To lower the transfer board

z

„

To raise the transfer board

To raise the transfer board

1

 Slowly lift the transfer board, which has been 

lowered, in the direction shown in the figure to 

the left. When it reaches the upper limit, the lug 
lowers and the stopper works.

2

 Confirm that the stopper is properly working 

(green shows in the indicator window).

3

 Press in the safety lock in the direction shown by 

the arrow 

 (transfer board side) until the red 

label can no longer be seen for proper locking.

NOTE!

z

z

Make sure not to let sheets get 
caught in the stopper or safety lock 
as it will hinder the functions of the 
parts.

*  In this manual, raising the transfer board entails 

completing all of the steps outlined in 

1

 to 

3

.

To lower the transfer board

1

 Release the safety lock.

2

 While lifting the lug, slowly lower the transfer 

board.

Lug

Red label

Safety lock

Lock

Stopper 
receptacle

Confirm that 

the indicator 
window 
shows green

Transfer board Warning label

Lift

Lug

Safety lock

Unlock

Stopper 
receptacle

Transfer board Warning label

Summary of Contents for KK-728EG

Page 1: ...Serious incident means any incident that directly or indirectly led might have led or might lead to any of the following a the death of a patient user or other person b the temporary or permanent ser...

Page 2: ...wheel system 11 3 Height adjustment 12 4 Raising the backrest 13 5 Transfer board 14 6 Restraint belt 16 7 Oxygen cylinder holder 17 8 IV Pole C57E 19 9 Accessory rail 21 10 Transfer mattress 21 6 PAT...

Page 3: ...ction base Accessory rail Oxygen cylinder holder Caster lock pedals Tray Base frame Height adjusting handle Main frame Leg section base IV Pole attachment hole Transfer board Restraint belt Transfer m...

Page 4: ...es IV Pole C57E Load capacity 1kg hook Oxygen cylinder holder Load capacity 15kg Tray Polypropylene PP Load capacity 10kg Hook 3K003355 4 pieces Load capacity 3kg piece Patient target groups Patient a...

Page 5: ...n could cause death or serious injury broken bones pressure or paralysis CAUTION Indicates that improper handling by ignoring this indication could cause light personal injury bruises wounds or cuts o...

Page 6: ...on the stretcher z z There is a risk of injury caused by falling or toppling from the stretcher In particular be careful with children Do not crawl under or place your head hands or legs under the st...

Page 7: ...rely secure the patient if using it with the two transfer boards locked in the horizontal level position z z Otherwise the patient could lose balance and fall out of the stretcher or fall over resulti...

Page 8: ...e stretcher Otherwise the stretcher may be damaged or deformed or even catch fire Use care when handling the gas spring z z Gas spring is used for the backrest lever If you attempt to disassemble or i...

Page 9: ...her is not in use store it with the casters facing in z z Otherwise people may trip over a caster and fall down or get injured Do not place objects weighing 10kg or more on the tray z z The tray could...

Page 10: ...ve amounts of water vapor or oil vapor z z Locations that tend to be exposed to high temperatures or humidity low temperatures or dry conditions Note The optimum ambient conditions for the stretcher a...

Page 11: ...operating only one caster lock pedal Caster lock pedals are located at the four corners of the stretcher allowing them to be operated at each side of head and foot ends To lock the casters z z Depress...

Page 12: ...stretcher may tip over resulting in an injury When the center wheel is OFF z z Depressing the center wheel locking pedal sets the center wheel to OFF to allow the stretcher to move in any direction in...

Page 13: ...heavy when it is turned close to the limit Do not turn the handle over the operating range Turning the handle forcibly over the range may result in damage or malfunction z z Do not put a leg on the ha...

Page 14: ...equipment Grip Grip Do not hold any other place besides the grip To raise z z Hold the grip of the back section base with one hand hold the backrest lever on the back section base with your other han...

Page 15: ...ay fall off and get injured To lower the transfer board To raise the transfer board To raise the transfer board 1 Slowly lift the transfer board which has been lowered in the direction shown in the fi...

Page 16: ...wly lift the transfer board to a horizontal level position See To raise the transfer board in p 14 2 Move the transfer board in the direction shown by the arrow while keeping it in a horizontal level...

Page 17: ...ay fall off or the stretcher may tip over resulting in an injury z z After placing a patient on the base back section base and leg section base of the stretcher use the two pairs of restraint belts to...

Page 18: ...nd flowmeter may be deformed resulting in injury 7 Oxygen cylinder holder The oxygen cylinder holder can contain a 500L 10cm to 11 5cm diameter oxygen cylinder The oxygen cylinder holder can be set ei...

Page 19: ...therwise the oxygen cylinder holder and flowmeter may be deformed resulting in injury Rotation stopper fitting Stopper Attachment part for horizontal setting Boss Thumb screw knob for locking a cylind...

Page 20: ...g the sliding pipe of the IV Pole adjust the sliding pipe length and then lock the sliding pipe by tightening the loosened thumb screw knob WARNING z z When loosening the thumb screw knob firmly grip...

Page 21: ...sliding pipe is loosened the sliding pipe may unintentionally fall resulting in injury z z Before use make sure the thumb screw knob for locking the sliding pipe is firmly tightened The sliding pipe m...

Page 22: ...the hook or accessory rail resulting in injury 10 Transfer mattress A patient can be transferred to from a bed without being lifted For the procedures see 6 PATIENT TRANSFER on p 22 to 25 The mattress...

Page 23: ...y or the side rail of the bed or the transfer board may get damaged or deformed Transfer board Transfer board Transfer board Transfer board receptacle When the transfer board is up When the transfer b...

Page 24: ...their hands together on their abdomen and pull out the side part of the transfer mattress that was folded in step 4 Grab handles of the transfer mattress handles on the stretcher side and pull the mat...

Page 25: ...ck pedal green to unlock the casters move the stretcher slightly away from the bed and raise the transfer board see p 14 WARNING z z When operating the transfer board place the patient s hands togethe...

Page 26: ...they will not hinder the transfer procedure Grab the handles of the transfer mattress and transfer the patient onto the bed by slowly pulling the mattress 4 After transferring the patient 1 Fold the s...

Page 27: ...as a support to stand up z z Do not transport a patient with the transfer board locked in the horizontal level position z z When transporting a patient be sure to fasten the restraint belts z z Operat...

Page 28: ...bnormal noise If water gets on them immediately wipe it off Maintenance and inspection items to be completed by the user of the product Main Unit The position operations using the height adjusting han...

Page 29: ...nium chloride e g Osvan Up to 0 2 Benzethonium chloride e g Hiamine Up to 0 05 Chlorhexidine gluconate e g Hibitane Up to 0 05 Sodium hypochlorite e g Milton It is highly recommended that you thorough...

Page 30: ...e center wheel is set to ON z z Turn OFF the center wheel 11 The transfer board cannot be lowered z z Check if the safety lock is ON z z Release the safety lock 14 15 The transfer board rattles z z Ch...

Page 31: ...od Paramount Bed keeps replacement parts necessary to maintain stretcher performance in stock for at least 8 years after the discontinuation of production 3 Questions regarding after sales service Con...

Page 32: ...Translation of the original Instruction Made by Torindo Co Ltd 2022 04...