Operations Manual: HPAC-100/125/150/200/250-RM SSPA
3 Operation of Stand-Alone Unit
Unit Status
Response to a Report Unit Status command. The Data field contains the status of
essential operating parameters for a particular unit.
Type:
RESPONSE
Pkt Length:
15
Destination:
M&C Address
Source:
SSPA System Address
Command:
0x92
Data:
Unit ID:
(1 byte, integer)
[ 1 | 2 | 3 ]
Control Location
(1 byte, integer)
0 = Local
1 = Remote
Configuration Mode
(1 byte, integer)
0 = Operation
1 = Setup
RF Muting
(1 byte, integer)
0 = OFF
1 = ON
Attenuation
(1 byte, integer)
[ 0 .. 20 ]
dB
Power
(2 byte, binary coded decimal)
[ 0.0 .. 55.0 ]
+dBm
Temperature
(1 byte, 2’s complement integer)
[ - 99 .. + 99 ]
°
C
Aux Switch Position
(1 byte, integer)
0 = NONE
No auxiliary switch
1 = POS 1
Auxiliary switch in position 1
2 = POS 2
Auxiliary switch in position 2
3 = BOTH
Auxiliary switch is stuck between positions
Unit Summary Fault
(1 byte, integer)
0 = OK
No summary fault.
1 = FAULT
Summary fault exists.
Length:
17 bytes