Instruction Manual ECO LIFT
Page 19 of 32
Version 2.03
Out2
Deactivated
Alm
Alarm; potentially occurring error message
Buttons
Circle icon
Menu; The user must not use this key as otherwise important
settings may be altered. Any modifications to the basic settings can
be retrieved from the internal memory in the event of a customer
complaint.
Arrow key
This key is used to increase the temperature.
Arrow key
This key is used to reduce the selected temperature.
R
Reset; If the "Alm" LED lights up, the alarm may be cleared using
this key. In case the alarm persists (LED does not turn off), please
contact your authorized distributor.
Adjusting the maximum temperature and defining the temperature unit
For setting a new maximum temperature as well as changing the applicable temperature
unit, please follow the steps described below in detail. In this context, also consider the
photo documentation provided hereafter for a better overview.
1.
Press and hold the "Circle" key on the left of the display for a few seconds. As a
result, " SET" should appear in the red text area of the display.
2.
Press the "Circle" key on the left of the display three times so that "UNIT" is now
shown in the red text area of the display.
3.
If required, the arrow keys on the display can be used to select a temperature
measurement, e.g. in Fahrenheit. The temperature unit currently active should also
be indicated by the green lettering on the display.
4.
Next, press the "Circle" key on the left of the display three times to adjust the
maximum temperature, until " SP1H" is visible in the red text area of the display.
1
2
3
Hold for several
seconds
Press 3x
Select with
arrow keys