Two different temperature values are shown on the display
Temperature green (see figure above)
This temperature is the set value currently determined by the user and therefore the desired
target temperature on the surface of the stainless steel top. It is possible to regulate this
temperature in steps of 0.1°C.
Attention:
The temperature of the heating top is measured with a so-called BI metal on
the underneath of the stainless steel plate. A deviation/spread of the
temperature between the surface and the bottom side of approx. 0.5 to 1°C
has to be anticipated. Heating is performed by means of an electric heating
mat. Since the system is rather idle and tends to reheat, one may notice that
the set temperature is slightly exceeded, even though the control of the
heating has already stopped operation. There may be a follow-up or reheating
of up to 3-5°C (depending on the environmental conditions). The longer the
system is in operation, the smaller the spread or overheating becomes, since
the heating control only slightly reheats when the temperature is just barely
below the specified target temperature.
Temperature red (see figure above)
This represents the temperature currently on the table top respectively the temperature of
the table top measured by the BI metal. Hence, the measuring point in that context is located
precisely in the center of the plate.
Additionally, there are further LEDs and buttons on the display whose function, operation
and purpose are explained in more detail below.
LEDs
Out1
This LED lights up when the heating has been switched on or when
the heating is activated.
Out1 lights up
→
system is
in heating mode
Current actual value
(temperature red)
Set point
(temperature green)
Temperature setting
(arrow keys)
On-Off-