Instruction Manual ECO LIFT 

 

Page 1 of 32  

Version 2.03 

 
 
 
 
 
 
 
 
 
 

 
 
 
 

 

 
 
 
 

ECO LIFT 

 
 
 
 
 
 

INSTRUCTION MANUAL 

 
 
 
 
 
 
 
 
 

 

 
 

For optimal and safe handling of the product, thoroughly reading and following the 

present Instruction Manual is essential. 

 

Summary of Contents for ECO LIFT 6H1.6P0.CCCC

Page 1: ...struction Manual ECO LIFT Page 1 of 32 Version 2 03 ECO LIFT INSTRUCTION MANUAL For optimal and safe handling of the product thoroughly reading and following the present Instruction Manual is essentia...

Page 2: ...nfortunately not offered for the EMEA and APAC regions Congratulations on your purchase of the present veterinary treatment and surgery equipment We are very delighted that you have made the decision...

Page 3: ...nt hydraulic version 13 Maneuverability optional castors as accessory 14 Adjustment of tilting mechanism 15 Adjustment of V Top mechanism for version with V Top 16 Version with heated top 17 Usage of...

Page 4: ...optimally and safely Hence panno med cannot be held liable for any damage or injury resulting from improper or incorrect use of the product The maintenance and care procedures for the product are desi...

Page 5: ...ibution Maximum longitudinal adjustment 80 Maximum V Top adjustment ca 45 per side V Top total ca 90 Net weight ca 54 72 kg 119 159 lbs Gross weight incl packaging ca 72 90 kg 159 198 lbs Weight depen...

Page 6: ...n the following equipment must be made available as well as prepared in order to be able to execute the described steps Allen key 5mm should already be contained in the scope of delivery Allen key 6mm...

Page 7: ...f 32 Version 2 03 Subsequently the rods are also threaded through at the passages of the respective clamping lever Finally the supplied plastic buffers are placed on the rods and screwed in place with...

Page 8: ...ecessary to screw the tube holder to the center strand of the V Top frame at the points shown above For this purpose 2 pcs of M6 screws and the corresponding washers intermediate layer are used Please...

Page 9: ...n screwed on condition These must be unscrewed from the table top accordingly before carrying out the assembly steps described in this context To stabilize and fix the rails tighten the screws accordi...

Page 10: ...nd secured in order to provide convenient access to the screw points of the table for the following assembly steps Hence the table top is required to be initially placed upon the column as well as its...

Page 11: ...t Lastly to stabilize and fix the entire table top and base connection tighten the mounted screws accordingly so that no more shaking of the parts is possible However please also make sure not to tigh...

Page 12: ...ber legs were previously unscrewed using the enclosed Allen key see picture documentation below Please be careful not to overtighten the screws as otherwise the screws or the table may be damaged Make...

Page 13: ...ressing the left button on the foot switch raises the table or initiates an upward movement In the opposite case pressing the right button lowers the table or initiates a downward movement The button...

Page 14: ...install retroactively optimal maneuverability of the table especially in narrow and angular environments should be ensured Each castor can be locked individually and separately For this purpose the le...

Page 15: ...such that the table top can now be maneuvered or freely adjusted within the tilting movement When the clamping lever is loosened the operator is able to move the plate to the desired tilting or horizo...

Page 16: ...pective table top half as well as for the V top adjustment and explicitly NOT the clamping lever described in the previous chapter for the adjustment tilt of the entire table top Consequently if the p...

Page 17: ...e clamping lever to the vertical position for locking the table top half at the new position Version with heated top The veterinary treatment and surgery table ECO LIFT with heatable stainless steel t...

Page 18: ...the control of the heating has already stopped operation There may be a follow up or reheating of up to 3 5 C depending on the environmental conditions The longer the system is in operation the smalle...

Page 19: ...mum temperature as well as changing the applicable temperature unit please follow the steps described below in detail In this context also consider the photo documentation provided hereafter for a bet...

Page 20: ...emperature unit and maximum temperature should already have been considered and accepted by the heating unit Usage of tie down cleat accessory part In case any tie down cleats or similar accessories h...

Page 21: ...xture provided for this purpose on the front side of the product as shown in the illustration to the right Removing the drain gutter from the fixture can also be done manually by pulling it out and li...

Page 22: ...14 of the product in question which is located on the type plate Potential error Recommended course of action At least one electric actuator does not work o Check if the battery if applicable is inse...

Page 23: ...ke sure to always read the product information and observe the instructions for the respective cleaning products and disinfectants The manufacturer cannot be held liable for damage injuries and impair...

Page 24: ...The use of additional protective coating in the form of varnish or paint is also strongly discouraged Engravings of names or other carved information on the stainless steel plate will immediately lead...

Page 25: ...cleaning recommendations for powder coated surfaces To ensure a long lifespan for your product the use of proper cleaning products for powder coated materials is strongly recommended Follow the instru...

Page 26: ...Instruction Manual ECO LIFT Page 26 of 32 Version 2 03...

Page 27: ...Instruction Manual ECO LIFT Page 27 of 32 Version 2 03...

Page 28: ...aspects 1 Comprehensive visual and functional test a Visual inspection All welding seams for cracks Surface damages Damages to the table top All cables bruises abrasions exposed wiring Visual inspect...

Page 29: ...product without prior permission from the manufacturer Any modification may result in unpredictable and unanticipated malfunctions and thus in damage or injury to the patient The manufacturer is not l...

Page 30: ...ess steel dish with grad 602 607 OP rail extension 200 mm pair 602 610 Instrument tray 300x200 mm 602 611 Arm rest for ophthalmology 602 614 Top extension 500x300 mm 602 615 Top extension 300x300 mm 6...

Page 31: ...ation of the device 5 The product may only be operated by properly instructed personnel 6 Any adjustment of the product may not be performed by the patient 7 Users are responsible for ensuring that th...

Page 32: ...tial to ensure that loads are distributed evenly over the table surface at all times High point loads must be strictly avoided Product warranty panno med provides a warranty for the proper functioning...