Aicure ANUP5255V2
Troubleshooting
70
14.2
Error message (Red-backlit screen and normal screen are
displayed alternately)
Message
Error description
Remedy
When UV control is ON and
the intensity control rate is
20% or more, the UV
intensity changes by
±
5% of
the set value or more.
UV intensity error
When the intensity control
rate is less than 20%, the UV
intensity changes by
±
7% of
the set value or more.
(1) Conduct UV
intensity check.
(See Page 41.)
(2) Replace the lamp.
Cover open
The left side cover is
disconnected or the screws
are loose.
Attach the left side
cover.
The lamp is not cooled
sufficiently.
Lamp illumination error
The lamp was forcefully
turned off when the cover
was removed or for another
reason.
After turning the “LAMP
ON/OFF” switch to OFF
and “POWER” switch to
OFF, turn the “POWER”
to ON again, and leave
until the cooling fan
stops (for 10 minutes).
Overheat
The temperature inside the
body rises abnormally high.
With the “POWER”
switch ON, turn the
“LAMP ON/OFF” switch
to OFF and leave for 10
minutes. (Caution)
Shutter error
Irradiation shutter failure
Contact the store from
which you made
purchase.
The cumulative lamp OFF
time is approaching the
lamp-change time. (30 hours
before)
Replace the lamp. (See
Page 61.)
Lamp-change indication
UV control
ON Replace
UV intensity 0000
Intensity rate 000 setting
0000
Backlight
Green
(ON)
* The “Replace” is displayed in
reverse video at the top right-hand
corner of the display screen.
When UV control is ON, the
intensity control rate is 98%
or more.
Use with a lower
intensity control rate.
If a necessary UV
intensity cannot be
obtained, the expiration
of the lamp’s useful life
is assumed. Replace
the lamp.
When the thermal relay is restored to normal condition after cooling, the error is reset.
Caution: