2.2
Installing T-Series Membrane Cassettes with Omega Membrane into Autotorque (AT)
and Manual Torque (MT) Holders
2.2.1
Preparation for Loading Cassettes
Procedure
1. Remove the retaining nuts, spacers and washers. Separate the sliding end plate
from the fixed flow-distribution manifold:
(i) for horizontal units, lift the end plate off the manifold;
(ii) for vertical units, slide the end plate away from the manifold about 15 to 22 cm
(6 to 9 in.).
2. Inspect and, if necessary, clean the cassette holder sealing surfaces. Inspect the
cassettes and gaskets prior to installation for any damage or foreign material
which could hamper sealing of the cassettes.
Figure 4
T-Series TFF Cassettes with Omega Membrane Product Packaging
2.2.2
Unpacking the Cassettes
Warning: This product is stored in 0.3 N Sodium Hydroxide. Refer to the MSDS for
proper protection and handling instructions.
Open the cassette package carefully!
To avoid accidental exposure to the storage reagent, open the cassette package
over a container or sink. Please see photos below for proper handling:
1. Remove cassette package from shipping container.
Figure 5
Cassette Package in Shipping Container
11