Fault Symptom
Possible Fault(s)
Control Panel inoperative
– Windows 95 does not recognise DSP assembly
– Applications software corrupted
– Control panel has become disconnected
– Poor connections via connections box
– Control panel is faulty
– DSP assembly faulty
Headbox controls do not work
– Headbox has become reset
No patient signals displayed
– Windows 95 does not recognise DSP assembly
– Applications software corrupted
– Headbox faulty (try with Amp Cal signal)
– Headbox lead faulty
– DSP assembly faulty
No Amp Cal signal displayed
– Applications software corrupted
– Headbox faulty (try with patient signal)
– Headbox lead faulty
– DSP assembly faulty
Unable to trigger signals
– Applications software corrupted
– Control panel faulty (try triggering using headbox controls)
_ Poor connections via connections box
– Headbox faulty (try triggering using control panel controls)
– DSP assembly faulty
No audio signal to loudspeaker
– Applications software corrupted
– Loudspeaker disconnected or faulty (try another loudspeaker)
– Control panel faulty
– DSP assembly faulty
No EMG Stimulus output
– Windows 95 does not recognise DSP assembly
– Applications software corrupted
– EMG Stimulus circuit in control panel or stim probe faulty
– Tower Internal cables, connections via I/O board and
connections box faulty
– DSP assembly faulty
Displayed EMG stimulus level
does not change as control adjusted
– Applications software corrupted
– Control board faulty (only applies to EMG1 stimulator)
– Stimulus probe faulty
– DSP assembly faulty
Incorrect level of audio output or
no audio output from EP Stimulator
board
– Windows 95 does not recognise EP Stimulator board
– Applications software corrupted
– EP Stimulator board faulty
– Calibration parameters stored on EP Stimulator board corrupted
– Internal cable disconnected or faulty
– Headphones faulty or connections via connector unit faulty
– DSP assembly faulty
No visual output from EP
Stimulator board
– Windows 95does not recognise EP Stimulator board
– Applications software corrupted
– EP Stimulator board faulty
– Monitor connections via connector unit faulty
– Monitor faulty
– DSP assembly faulty
Issue A May 1998
FAULT FINDING
Chapter 8 Page 2
Synergy Mobile Workshop Manual