INSTALLATION AND OPERATION
Issue A May 1998
Synergy Mobile Workshop Manual
Chapter 3 Page 5
Example B - Both Synergy and the other electrical equipment are connected to the local protective
earth in their respective rooms. As the may be a potential difference between the protective earths
in the two rooms Synergy requires a second separate protective earth connection in the medically
used room to prevent allowable leakage currents being exceeded in the event of failure of the first
Portable multiple socket outlets
be used.
must not
protective earth to Synergy.
NO !
Synergy mains
cable including
protective earth
Other equipment mains
cable including protective
earth
Signal
cable
Example B
MEDICALLY USED ROOM
Synergy
Wall outlet
socket
PC tower
Printer
Monitor
Isolating
transformer
Mains
inlet
Isolated mains
outlets
Other electrical
equipment
OTHER ROOM
Signal
cable
Wall outlet
sockets
Additional protective
earth conductor
connected between
the metalwork of the
Synergy mains
isolating transformer
module and the earth
pin of a
wall socket in the
medically used room
separate
Additional protective
earth point