8.7
Exposure with the wireless X-ray button (optional)
It is possible to make an exposure using the wireless X-ray button.
Proceed as follows:
1.
Press and release the wireless X-ray button. The green LED will turn
on, indicating that the communication with the timer was successful.
2.
Move away the desired distance (no greater than 5m), in the opposite
direction of the X-ray beam.
3.
Press the X-ray emission button and hold it down during the entire
exposure.
The procedure continues as described in paragraph 8.6, points 4, 5
and 6.
4.
At the end of exposure, the green LED on wireless button flashes
quickly two times.
In the following table are described the different wireless X-ray button
green LED status.
Description
Flashing
frequence
Duration
Action
Ready to exposure
Fixed
Fixed
--
Wireless button
communication failure
High
2 seconds
Check that the device is
not switched off or in
sleeping/cooling mode.
If the device is active,
perform matching
procedure (see paragraph
Matching procedure
failed
High
10 flashes
Repeat matching
procedure (see paragraph
Matching procedure OK Low
10 flashes
--
Low battery
Very high
2 sequences
of 5 flashes
Replace battery (see
paragraph 8.7.1)
Very low battery
Very high
4 sequences
of 5 flashes
Replace battery (see
paragraph 8.7.1)
Wireless button faulty
LED off
--
Contact the technical
support
Summary of Contents for Ow-RX
Page 1: ...Version July 2019 Rev 7 Ow RX User Manual 0051...
Page 2: ......
Page 4: ...USER MANUAL Revision history Ow RX Rev 6 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 22: ...USER MANUAL Description Ow RX Rev 6 16 5 DESCRIPTION 5 1 Identification plates 5 1 2 6 4 3...
Page 73: ...USER MANUAL Maintenance Rev 6 Ow RX 67...
Page 74: ...Cod 6959902203_Rev 6 0051...