3.1.1
Electromagnetic emissions
In accordance with the IEC 60601-1-2 standard, Ow-RX is suitable for
use in the electromagnetic environment specified below.
The customer or user of the system must ensure that it is used in the
said environment.
Emissions test
Conformity
EMC environment of use
RF emissions
CISPR 11
Class B
Ow-RX is suitable for use in all domestic
environments and in environments
directly connected to the mains power
supply at low voltage that supplies
buildings for domestic use.
Group I
Ow-RX uses RF power only for its
internal functioning. As a result, its RF
emissions are very low and most likely
will not cause any interference in
electronic devices located nearby.
Harmonic emissions
IEC 61000-3-2
Class A
Flicker/voltage
fluctuation emissions
IEC 61000-3-3
In
compliance