![Owandy Radiology Ow-RX User Manual Download Page 32](http://html1.mh-extra.com/html/owandy-radiology/ow-rx/ow-rx_user-manual_1663721032.webp)
USER MANUAL
Technical data
Ow-RX
(Rev. 6)
26
6.
TECHNICAL DATA
Technical characteristics
Equipment
Ow-RX
Manufacturer
OWANDY RADIOLOGY
77183 Croissy-Beaubourg - France
Class
Class I with type B applied parts
(according to EN 60601-1
classification)
Protection degree
IPX0 standard device
Line voltage
99-264 V
∼
Rated voltage
110-240 V
∼
Line frequency
50 / 60 Hz
Line current
5.4 A rms impulsive @ 110 V
∼
2.4 A rms impulsive @ 240 V
∼
Maximum line current
7 A rms impulsive @ 99 V
∼
Technical factors for maximum line
current
70kV, 6mA
Absorbed power
582 VA @ 110 V
∼
570 VA @ 240 V
∼
Maximum apparent line resistance
0.4
Ω
(99-180 V
∼
)
0.8
Ω
(181-264 V
∼
)
Line voltage regulation
≤ 3%
Mains fuse (F1)
T 6.3 A - 250 V
Mains fuse (F3 - only for mobile version)
T 6.3 A - 250 V
Selectable times
from 0.01 to 2.00 s in 36 steps
Automatic selection
882 pre-programmed times (7 anatomic -
3 sizes - 3kV - 2 SID- 3 receptors)
Time accuracy
±
5 %
±
2 ms
High voltage values
60-65-70 kVp selectable
Tubehead current
6 mA
kV accuracy
±
8 % @ rated voltage
Tubehead anodic current accuracy
±
10 % @ rated voltage
Maximum exposure time
2.0 s
Timer size
284
×
253
×
123.3 mm
Summary of Contents for Ow-RX
Page 1: ...Version July 2019 Rev 7 Ow RX User Manual 0051...
Page 2: ......
Page 4: ...USER MANUAL Revision history Ow RX Rev 6 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 22: ...USER MANUAL Description Ow RX Rev 6 16 5 DESCRIPTION 5 1 Identification plates 5 1 2 6 4 3...
Page 73: ...USER MANUAL Maintenance Rev 6 Ow RX 67...
Page 74: ...Cod 6959902203_Rev 6 0051...