USER MANUAL
Error messages on the display
(Rev. 6)
Ow-RX
61
9.3
NON fatal errors
Situations that do not directly involve the safety of the operator, the
patient or the system are considered as resettable anomalies.
The error condition prevents additional exposures until it is reset
by pressing the "Increase" key.
Displayed
message
Type of ABNORMALITY
ACOUSTIC
signal
CH0
Memory checksum error (EEPROM)
None
E11
X-ray button active after the cooling phase
None
E12
Release of X-ray button during emission
None
E13
Release of button during the pre-heating
phase
None
WARNING:
If CH0 is displayed, the EEPROM will be reinitialized to the default
parameters. In case of errors at switch ON or setup messages
different than the original ones, call the technical service to restore
customizations.
If E12 is displayed, the X-ray button was released with the emission
in progress, therefore the film must be replaced or the image
receptor must be restarted to obtain the diagnostic results.
If E11 is displayed, release the X-ray button; if this has not been
pressed, this indicates a fault, therefore call the support service.